TCP4 converts thermo-EDN from thermocouples into standard current or voltage signal.
TCP4 has built-in electronic compensation of the temperature of the cold end for the selected type of thermocouple. Number of channels in one device - 4. Type of thermocouples: E, J, K, R, S, T. Measured temperature range - according to order. Maximum summary error: £ ±(0.5 °C+0.2 % of the full temperature range). Dimensions with cable glades: 180х135х40 mm. Mounting on the wall. Variants: TCP4 Output signal: 4-20 mA. Power supply: 24 VDC. Consumption: £ 3.5 W. TCP4-1 Output signal: 0-10 VDC. Power supply: 24 VDC. Consumption: £ 0.5 W. TCP4-2 Output signal: 0-10 VDC. Power supply: 12 VDC. Consumption: £ 0.5 W. TCP4-3 Output signal: 0-5 VDC. Power supply: 24 VDC. Consumption: £ 0.5 W.
|